Device information, Calibration/adjustment, Weighing – Rice Lake Secura - Quintix and Practum Series - User Manual User Manual

Page 26

Advertising
background image

26

System Settings (Setup Menu)

User Manual for Secura, Quintix, Practum

Device

Information

Device information

Manufacturer: Sartorius
Model: SECURA2102
Serial number:

0012345678

Version BAC:

00-50-02.xx CN:1701

Version APC:

01-70-02.xx CN:6587

The manufacturer, model, serial number, and software version of your balance are

displayed here.

Calibration/Adjustment

Calibration/Adjustment

isoCAL

Calibration report

With

Secura

®

and Quintix

®

models, saved reports about the most recent calibration

procedures, including the identified deviations, can be viewed and output to a PC
or Sartorius lab printer.
With Secura

®

models it is also possible to set how the automatic calibration/

adjustment function isoCAL is started.

isoCAL

menu option (Secura

®

only):

t

Select the desired start option for isoCAL.

Off

: isoCAL is switched off.

Info, manual start

: A note appears on the display when predefined time

intervals or temperature values are exceeded. isoCAL can be started
manually.

Automatic

: isoCAL is switched on. Calibration/adjustment starts

automatically when predefined time intervals or temperature values are
exceeded.

Calibration report

menu option (Secura

®

and Quintix

®

only):

t

Select the desired report by date.

If there is more than one calibration report per date, use the

<

and

>

buttons to

navigate through the reports.

t

Select

a

to print the displayed report.

Weighing

Weighing

SQmin

Application

Stability signal

Ambient conditions

Safety level

A number of basic settings for the weighing functions can be defined here. Options

are limited for verified balances. All settings given in the menu are permitted.

SQmin

menu option (Secura

®

only):

If the SQmin function is set up on the balance, the function can be switched

On

or

Off

here (see “SQmin Minimum Weighing (Secura

®

only)" on page 37).

Safety level

menu option (Secura

®

only):

t

Select the desired safety level for

SQmin

,

isoCAL

, and

LEVEL

Low

: Information about display elements given in red. Weighing is still

possible.

Standard

: As with

Low

, but the weight value is also displayed in gray and

the warning symbol is shown. The weight values are marked as not safe.
A warning appears after 60 seconds if the situation is ignored.

High

(factory setting): As with

Standard

, but the weight value in the

applications cannot be printed out or saved if the value is smaller than the
specified minimum initial weight (SQmin).

Calibration report

2012-10-04: 2 Protocols

2012-10-01: 2 Protocols

2012-09-25: 3 Protocols

2012-09-24: 3 Protocols

2012 09 20: 4 Protocols