Rice Lake Secura - Quintix and Practum Series - User Manual User Manual

Page 49

Advertising
background image

49

Density

User Manual for Secura, Quintix, Practum

Density of liquid

g/cm

3

0.9982

t

To change the density of the buoyancy liquid, select the gray button.

Density

Density of liquid

Accuracy

0.9982

0.9980

0.9978

0

0.0

0.00

0.000

y The

Density

settings window appears.

t

Enter the density of the buoyancy liquid on the left.

Select a value or select

t

, enter the desired value, and confirm with

l

.

t

Select the accuracy of the density result on the right.

t

Select

l

to confirm.

t

Attach the sample holder to the frame of the density determination kit.

t

Fill the beaker in the density determination kit with the buoyancy liquid.

Ensure that the sample holder is sufficiently immersed in the liquid in order to
hold the sample later.

t

If distilled water is used as the buoyancy liquid, add three drops of tenside to

reduce the effect of surface tension on the measurement result.

t

Select

D

to tare.

Net

Max 220 g

d=0.0001 g

0.0000

Density of liquid

g/cm

3

0.9982

y The balance is tared with the prepared density determination kit.

t

To begin density determination, touch

w

.

t

Place the sample on the weighing pan above on the frame of the density

determination kit.

Max 220 g

d=0.0001 g

4.8909

Weight in air

END

+

Net

y The weighing display shows the step

Weight in air

.

Advertising