Rice Lake Secura - Quintix and Practum Series - User Manual User Manual
Page 49
49
Density
User Manual for Secura, Quintix, Practum
Density of liquid
g/cm
3
0.9982
t
To change the density of the buoyancy liquid, select the gray button.
Density
Density of liquid
Accuracy
0.9982
0.9980
0.9978
0
0.0
0.00
0.000
y The
Density
settings window appears.
t
Enter the density of the buoyancy liquid on the left.
Select a value or select
t
, enter the desired value, and confirm with
l
.
t
Select the accuracy of the density result on the right.
t
Select
l
to confirm.
t
Attach the sample holder to the frame of the density determination kit.
t
Fill the beaker in the density determination kit with the buoyancy liquid.
Ensure that the sample holder is sufficiently immersed in the liquid in order to
hold the sample later.
t
If distilled water is used as the buoyancy liquid, add three drops of tenside to
reduce the effect of surface tension on the measurement result.
t
Select
D
to tare.
Net
Max 220 g
d=0.0001 g
0.0000
Density of liquid
g/cm
3
0.9982
y The balance is tared with the prepared density determination kit.
t
To begin density determination, touch
w
.
t
Place the sample on the weighing pan above on the frame of the density
determination kit.
Max 220 g
d=0.0001 g
4.8909
Weight in air
END
+
Net
y The weighing display shows the step
Weight in air
.