Rice Lake Secura - Quintix and Practum Series - User Manual User Manual
Page 68
68
Calibration and Adjustment
User Manual for Secura, Quintix, Practum
Calibration/Adjustment using an
External Calibration Weight
h
An external calibration weight is required for this function.
Please note the tolerance of the calibration weight being used.
In verified balances with accuracy class
K, external calibration/adjustment is
locked when the balances are used for legal metrology.
Secura
®
and Quintix
®
: In verified balances with accuracy class
k, external
adjustment can only be accessed when the access switch is open.
t
Make sure that the weighing pan is empty.
t
Select
A
to zero the balance.
t
Select the Menu key at the bottom left of the display in the current application.
Components
Statistics
Mixing
Weighing
y The menu appears.
t
Select
the
CAL
button to access the calibration function.
Calibrate balance
CAL-Extern
CAL-Intern
y The
Calibrate balance
window appears.
t
Select
CAL-Extern
.
y A message appears on the display, prompting the user to place the calibration
weight on the weighing pan.
t
Place the calibration weight on the weighing pan.
y Calibration begins automatically.
Date and Time:
14-Jun-2012 14:13
Calibration function:
External calibration
Start calibration:
Start: manually
Calibration weight:
Set 2000.00 g
Calibration deviation:
Dev 0.02 g
Adjustment function:
External adjustment
Adjustment deviation:
Dev 0.00 g
Report
y A report appears on the display (Secura
®
and Quintix
®
only).
The report indicates the deviations identified during calibration. The report also
gives the result of the adjustment process.
t
Select
<
.
y The balance is now calibrated and adjusted.