Calibration and Adjustment

User Manual for Secura, Quintix, Practum

Calibration/Adjustment using an
External Calibration Weight

h

An external calibration weight is required for this function.

Please note the tolerance of the calibration weight being used.

In verified balances with accuracy class

K, external calibration/adjustment is

locked when the balances are used for legal metrology.
Secura

®

and Quintix

®

: In verified balances with accuracy class

k, external

adjustment can only be accessed when the access switch is open.

t

Make sure that the weighing pan is empty.

t

Select

A

to zero the balance.

t

Select the Menu key at the bottom left of the display in the current application.

Components

Statistics

Mixing

Weighing

y The menu appears.

t

Select

the

CAL

button to access the calibration function.

Calibrate balance

CAL-Extern

CAL-Intern

y The

Calibrate balance

window appears.

t

Select

CAL-Extern

.

y A message appears on the display, prompting the user to place the calibration

weight on the weighing pan.

t

Place the calibration weight on the weighing pan.

y Calibration begins automatically.

Date and Time:

14-Jun-2012 14:13

Calibration function:

External calibration

Start calibration:

Start: manually

Calibration weight:

Set 2000.00 g

Calibration deviation:

Dev 0.02 g

Adjustment function:

External adjustment

Adjustment deviation:

Dev 0.00 g

Report

y A report appears on the display (Secura

®

and Quintix

®

only).

The report indicates the deviations identified during calibration. The report also
gives the result of the adjustment process.

t

Select

<

.

y The balance is now calibrated and adjusted.