FT-11(D) Technical Manual, Rev. 1.35 November 2010

Page 17 of 52

6.3. Configuration

Parameters

[1--] Configuration Block

[11-] Start Up

[111 X] Auto Power Off

0 : Auto power off disabled (default)

1 : Power off after 3 minutes of stability or no key function

2 : Power off after 10 minutes of stability or no key function

[112 X] Store Tare at Power Off

0 : Disabled (factory default)

1 : Enabled

[113 X] Auto Clear Tare

0 : Disabled (factory default)

1 : The scale returns to gross mode after unloading

[115 ] Key lock (for FT-11 only)

[ ABCDE]

When the key lock function is activated, all key(s) set to “1” in this parameter are
locked.

A : On/Off

B : Setpoint

C : Zeroing

D : Taring / Clear

E : Print

[12-] Filter

In this block the proper filter values according to the operating conditions can be entered. One of the most
important features of the indicator is viewing the filter characteristics on the display and with the help of this
option, you can select the most suitable filter without leaving the setup mode.

[120 X ] Filter

For FT-11 the filter value can be set from 0 to 9, for FT-11D the value range is from 0 to 4. When you
enter this parameter and press the <Enter> key while [120 X] is shown on the display, the weight
variation can be seen on the display. The value of the filter can be changed by using the <

> and <>

keys and the weight variation for each value is shown on the display. After finalising the filter selection
you can go to the next step by pressing the <F> key.

[14-] Entries

In this block you can enter the initial CN (consecutive number).

[142 ] Label No. (CN)

[ XXXXXX ] The desired value is entered via <Tare> and <Zero> keys and saved by pressing

<Enter>. If the number exceeds 65 535 it will automatically restart at 1.

6.4. Scale

Parameters

[2--] Scale Block

The parameters related to the weighing operation are located here.

[20-] Set Up

[200 X] Approved

0 = No

1 = Yes

[201 X] Increased Indication

0 = by pressing key

1 = Always increased indication

[202 X] Power On Zero

This parameter setting controls if after power-on the scale will get automatically zeroed if the weight is in
the percentage of the zeroing range. If the weight is not in the zeroing range the display will show [E E E
] message until you press the <Enter>.

0 = disable

1 = ± 2%

2 = ± 10%

[203 X] Zeroing Range

0 = disable

1 = ± 2%

2 = ± 20%