Appendix 4: modbus rtu data structure – Flintec FT-11D User Manual

Page 42

Advertising
background image

FT-11(D) Technical Manual, Rev. 1.35 November 2010

Page 42 of 52

Appendix 4: Modbus RTU Data Structure

If the instrument is programmed for Modbus RTU, it can be used as a Modbus RTU slave on a RS485
communication line or on the Ethernet communication line. The Modbus slave adress is defined in the
parameter [Address]. Function code “0x03” and function code “0x10” are supported.

Read

command

Modbus

register

Function

code

Register

address

No. of

registers

Definition

Display value

40 001

03

00,00

00,02

Current display value; Gross or Net

D0

Busy

1 – System Busy

D1

Data ok

1 – Data ok
0 – Error (See D13…D15)

D2

Motion

1 – Weight is not stable

D3

Net Mode

1 – Net mode

D4

N/A

D5

Output 1

D6

Output 2

D7

Output 3

1 – Output active

D8

Input 1

D9

Input 2

D10

Input 3

1 – Input active

D11

D12

N/A

0 No

Errors

1

ADC out of range

2 ADC

overrange

3 ADC

underrange

4 System

fault

Status 40

003 03 00,02 00,01

D13
D14
D15

Transfer
binary (D13
to D15) to
decimal and
compare to
error code

5

FT not in weighing mode

Tare weight

40 004

03

00,03

00,02

Current tare weight

Gross Weight

40 006

03

00,05

00,02

Current gross weight

Setpoint 1

40 012

03

00,0B

00,02

Setpoint 1 buffer memory; replied value in hex

Setpoint 2

40 014

03

00,0D

00,02

Setpoint 2 buffer memory; replied value in hex

Setpoint 3

40 016

03

00,0F

00,02

Setpoint 3 buffer memory; replied value in hex

1

Ready for calibration

3

Zero calibration in process

4

Span calibration in process

9

Error (Refer to D8 ... D15 )

D0…D7
Calibration
Process
Status

1

Calibration Timeout –
Restart calibration

2

ADC Error
– Re-energize the instrument
– If seen again, change board.

3

Instrument cannot be calibrated –
Check load cell cable
– Re-energize the instrument

34

Instrument cannot be calibrated –
Load cell signal is too low or too
high

35

Calibration Error
– Calibration test weight is too
small, increase calibration weight
value (40031-32)
– Check load cell connections

37 Scale

unstable

– Wait until scale becomes stable
– Check grounding wiring

38

Calibration Jumper is not installed-
Check Calibration jumper

Calibration

Status

40 033

03

00,01

D8…D15
Calibration
Errors

1

Ready for calibration