FT-11(D) Technical Manual, Rev. 1.35 November 2010

Page 22 of 52

8. Communication

8.1. Overview about Communication

Application Hardware

channel

Operation

mode of the

interface

Data export to PC (non-approved);
Remote display;
Remote control with ASCII entries
(Z, T, C)*

Continuous output
mode;
see Chap. 8.2.3
and Appendix 2

IndFace-Software (FT-11 only);
Process control (PLC, DCS,
SCADA...)

Host mode (for
FT-11 only);
see Chap. 8.2.4 &
Appendix 3

Data export to PC (legal for trade
with alibi memory);
Output to printer;
Remote control with ASCII entries
(P, Z, T, C)*

Standard Serial Interface (RS232C)

OR

Option Serial Interface 2 (RS232C)

OR

Option Serial Interface 3 (RS232C, RS485, 20 mA TTY)

OR

Option Ethernet (for FT-11 only; see Chap. 8.3)

Print mode;
see Chap. 8.2.5

Option Serial Interface 2 (RS485)

OR

Option Ethernet (for FT-11 only; see Chap. 8.3)

Modbus RTU (for
FT-11 only);
see Appendix 4

Option Profibus (for FT-11 only; see Chap. 8.4 + Appendix 5) Profibus DP
Option ProfiNet (for FT-11 only; see Chap. 8.5 + Appendix 5) ProfiNet

Process control (PLC, DCS,
SCADA...)

Option CANopen (for FT-11 only, see Chap. 8.6 + App. 6)

CANopen

* Z = Zero; T = Tare; C = Clear; P = Print

8.2. Serial Interface and Printer

The instrument has a standard serial port for connecting peripheral equipment.
The serial port is suitable for bi-directional communication. If you transmit ASCII codes of P, Z, T or C letters to
the serial port, the indicator will behave like the related keys have been pressed. If optional serial interfaces are
installed, only continuous data output can be programmed for more than one interface (see chapter 8.2.7 for
setup details).

8.2.1.

Standard Serial Interface Connector (RS232C)

The instrument has a standard serial port for connecting to peripherals.

Standard Serial interface; Parameter group [00-]

Baudrate

1200, 2400, 4800, 9600, 19 200, 38 400 or 57 600 baud

Data format

8 bit no parity OR 7 bit even parity OR 7 bit odd parity

Start / Stop bit

1 start bit and 1stop bit


The connection of the standard RS232C of the instrument should be made as given in the table below.

Definition

Desk type housing, panel type

housing; Pin no. (D-Sub, 9-pin, male)

Stainless steel housing

Pin no. (J11 connector)

RS232C;

typical PC allocation

TXD 2

3

3

RXD 3

4

2

GND 7

2

5

Shield Connector

body

1


You have to connect each TXD (output) with RXD (input) at the other end.

8.2.2.

Optional Serial Interface Connector (RS232C, RS485, 20 mA TTY)

Additional serial interfaces can be added in addition to the standard RS232C port.

2

nd

Serial interface

Optional RS232C; Parameter group [01-]

3

nd

Serial interface

Optional RS232C / RS485 / 20 mA TTY current loop* (field changeable); Parameter group [02-]

* 20 mA TTY current loop not available with stainless steel housing


If parameter [012] is set to 2 (= hardware handshake), the second serial interface cannot be used as RS232C
port. Up to 32 instruments can be connected with enabled RS485 port.