FT-11(D) Technical Manual, Rev. 1.35 November 2010

Page 34 of 52

[132 X] Input 2

0 = Disabled (factory default)

1 = Zero

2 = Tare

3 = Clear

4 = Print

5 = Key lock

[133 X] Input 3

0 = Disabled (factory default)

1 = Zero

2 = Tare

3 = Clear

4 = Print

5 = Key lock

9.2.3. Entering

Setpoints

Press the <SP> key to enter the setpoint memory. The setpoint values will be defined by the <Zero> and
<Tare> keys and confirmed with the <Enter> key. You will advance to the next setpoint. Here you can select
between setpoints via the <

> and <> keys. Quit without saving by pressing the <F> key.

9.3. Binary Output Option

This option adds a 17-bit binary output to the instrument which changes in the value range between “0” and
“131 071” depending on the displayed value and the zeroing input status.
The value of the binary output will change related to the display value as stated in the table below.

BINARY

OUTPUT

Condition (On Displayed)

Data

Negative

Error

Over High

Low

High

Under High

High

High

ADC Out

High

Low

High

Weight indication

Binary

X

Low


The pin configuration of the D25 female binary connector is given below.

Pin no.

Connection

Definition

1

24V DC

POWER SUPPLY

14-2 GND

POWER

SUPPLY

15 TEST

INPUT

3 D0

2

0

16 D1

2

1

4 D2

2

2

17 D3

2

3

5 D4

2

4

18 D5

2

5

6 D6

2

6

19 D7

2

7

7 D8

2

8

20 D9

2

9

8 D10

2

10

21 D11

2

11

9 D12

2

12

22 D13

2

13

10 D14

2

14

23 D15

2

15

11 D16

2

16

24

Strobe

1 = Data Ready

12 Zero

Input

25 Error

Output

13 negative

Output

D25 body

Schirm

“Data Ready” means that the weight data can
be read from the output.
If the output value is low, a wrong reading of
the weight value is probable.
If you apply a test input, the weight data
output will be interrupted and all outputs will
be activated sequentially during the test input
is active.
Outputs are PNP open collector and common
is ground.
The binary output card has to be supplied with
24 V DC ( max. 28 V DC ), maximum current
consumption can be up to 200 mA.