Digital i/o option, ] configuration block, ] digital inputs / outputs – Flintec FT-11D User Manual

Page 33: Electrical connections, Setup, 130 x] outputs, 131 x] input 1

Advertising
background image

FT-11(D) Technical Manual, Rev. 1.35 November 2010

Page 33 of 52

9.2. Digital I/O Option

9.2.1. Electrical

Connections

It is possible to use control inputs and setpoint outputs if you add an optional digital I/O board (3 inputs and 3
outputs) to the instrument. The electrical connection of the inputs and outputs should be made as shown in Fig.
9.1.

Figure 9.1

Digital I/O ports connection diagram


Voltage Current

Digital input

20 to 28 V DC

Max. 30 mA

Digital output

20 to 28 V DC

Max. 300 mA (sum of all 3 outputs)


The power for the digital I/O has to be supplied from an external power supply module.


The pin layout of the parallel Input/Output connector is given below.

Pin no. for stainless steel housing

Definition

Pin no. for desk and panel types

(DB-Sub, 25-pin, female)

J10 connector (Option 1)

JR1 / JR2 (Option 2 or 3)

0 V

13

1

1

+24V

12 - 24 - 25

2

2

Output 1

1

7

8

Output 2

3

8

7

Output 3

5

9

6

Input 1

11

3

10

Input 2

10

4

11

Input 3

9

5

12

Shield

Connector Body

Not Used

Not Used

9.2.2. Setup

[1--] Configuration Block

In this block the parameters are located which are being used to set up the digital inputs and outputs.

[13-] Digital Inputs / Outputs

FT-11(D) has optionally 3 digital inputs and 3 digital outputs.

[130 X] Outputs

The outputs operate with respect to the displayed weight value. In the 3

rd

selection the Output 3 operates

as Error and Alarm output.

0 = Disabled (factory default)
1 = (Output1 = Sp1), (Output2 = Sp2), (Output3 = Sp3)
2 = (Output1 = Sp1), (Output2 = Sp2), (Output3 = Stable)
3 = (Output1 = Sp1), (Output2 = Sp2), (Output3 = Error and Alarm)

[131 X] Input 1

0 = Disabled (factory default)

1 = Zero

2 = Tare

3 = Clear

4 = Print

5 = Key lock

Warning:

This input will be used as tilt switch input for tilting applications (parameter [207]  0)

Advertising