Page 24: Setup, 000 x ] data format, 001 x] baud rate, 002 x] handshake, 003 xx] address, 004 x] data length and parity, 005 x] checksum

Advertising
background image

FT-11(D) Technical Manual, Rev. 1.35 November 2010

Page 24 of 52

2. Multi Line

Flintec GmbH

www.flintec.com

Germany

Flintec GmbH

www.flintec.com

Germany

By pressing the <Print> key you can send the data in multiple lines as shown
in the labels at the left.

3 lines of header and 1 line of footer can be downloaded to the instrument (in
host mode).

The data output structure is defined by the printer parameters.

3. Title and Footnote (for FT-11 only)

It is possible to load 3 lines of header and 1 line of footer to FT-11 and print a label with title and footnote in print
mode.
Title and footnote can be loaded to the indicator from a PC via the serial port in host mode. The FLINTEC
IndFace software can be used for this. The free of charge interface software can be downloaded from the
FLINTEC web site

www.flintec.com

.

8.2.6. Setup

[0--] Interface Block

You can access the parameters of the instruments’s serial interface(s) in parameter section [0--].
The data output modes can be used only once per instrument except for continuous data output.

[00-] Serial Interface 1 (Standard serial interface)

[000 X ] Data Format

0 : No data transfer

1 : Continuous data output

2 : Print mode ( refer to parameter [040] )

3 : Host mode

[001 X] Baud Rate

0 : 1 200 Baud

1 : 2 400 Baud

2 : 4 800 Baud

3 : 9 600 Baud

4 : 19 200 Baud

5 : 38 400 Baud

6 : 57 600 Baud

[002 X] Handshake

0 : No Handshake

1 : Xon/Xoff

[003 XX] Address

The address range is 1 to 99. If you enter 0, the indicator will operate without an address.

[004 X] Data Length and Parity

0 : 8 bit, no parity

1 : 7 bit, odd parity

2 : 7 bit, even parity

[005 X] Checksum

0 : Checksum byte disabled

1 : Checksum byte enabled

[01-] Serial Interface 2 (Optional serial interface)

[010 X ] Data Format

0 : No data transfer

1 : Continuous data output

2 : Print mode (refer to parameter [040])

3 : Host mode

[011 X] Baud Rate

0 : 1 200 Baud

1 : 2 400 Baud

2 : 4 800 Baud

3 : 9 600 Baud

4 : 19 200 Baud

5 : 38 400 Baud

6 : 57 600 Baud

[012 X] Handshake

0 : No Handshake

1 : Xon/Xoff

2 : Hardware

Warning:

If you choose hardware as handshake, the serial interface 3 can not be used as RS232C.

[013 XX] Address

The address range is 1 to 99. If you enter 0, the indicator will operate without an address.