FT-11(D) Technical Manual, Rev. 1.35 November 2010

Page 18 of 52

[204 X] Auto Zero Tracking

AZT automatically re-adjusts the scale to zero for compensating defined small deviations around the
centre of zero.

0 = disable

1 = ± 0,5e

2 = ± 1e

3 = ± 3e

[205 X] Tare

0 = Disabled

1 = Multi tare via key

2 = Tare via key in gross mode

[206 X] Motion Detector

This parameter defines the sensitivity level which will determine what is considered as stable.

0 = ± 0.3e

1= ± 0.5e

2= ± 1e

3= ± 2e

4 = No motion detector

[207 X] Tilt Switch (for FT-11 only)

The tilt switch is being used to prevent wrong weighing results in mobile scales due to the slope of the
floor. If this parameter is enabled the digital input 1 must be connected to the tilt switch.

0 = Disabled

1 = Normally Open contact

2 = Normally Closed contact

[208 X] Stability Time

This parameter defines the time period (up to 9.9 s) which will determine what is considered as stable.

[21-] Scale Build

The capacity and the resolution of the scale will be defined here.

[210 X] Scale Type

The capacity and resolution parameters will vary according to the selection here.

0 = Single Range

1 = Dual Range

You will reach the next parameter by pressing <Enter>. Please note that the following parameters will
vary according to this selection.

If single range has been selected:

[212 ] Capacity

Press the <Tare> key to reach this parameter.

[CAP ]

[ XXXXXX ]

The capacity of the scale will be entered here by <Tare> and <Zero> keys.
The value will be confirmed by pressing <Enter>.

[d ]

[ XXXXXX ]

Display resolution of the scale will be selected by <Zero> key.
The selection will be confirmed by pressing <Enter>.

If dual range has been selected:

[211 ] 1. Capacity

Press the <Tare> key to reach this parameter.

[CAP1 ]

[ XXXXXX ]

The capacity of the first weighing range (Select from Appendix 6) will be entered
here by <Tare> and <Zero> keys and confirmed with <Enter>.

[d1 ]

[ XXXXXX ]

Display resolution of the first weighing range will be selected by <Zero> key and
confirmed with <Enter>.

[212 ] 2. Capacity

Press the <Tare> key to reach this parameter.

[CAP2 ]

[ XXXXXX ]

The capacity of the second weighing range (Select from Appendix 7) will be entered
here by <Tare> and <Zero> keys and confirmed with <Enter>.

[d2 ]

[ XXXXXX ]

Display resolution of the second weighing range will be selected by <Zero> key and
confirmed with <Enter>.


Attention:

After the scale parameters have been changed the scale has to be calibrated.