FT-11(D) Technical Manual, Rev. 1.35 November 2010

Page 47 of 52

Appendix 6: CANopen Data Structure (for FT-11 only)

The output from FT-11 on PLC requests is in two double words and according to the table below.

FT-11 Output

Bit

no. 15 14 13 12 11 10 9 8 7 6 5 4 3 2 1

0

W31

MSB

W30 W29 W28 W27 W26 W25 W24 W23 W22 W21 W20 W19 W18 W17

W16

TxPDO 1
(T_DW1)

W15 W14 W13 W12 W11 W10 W9 W8 W7 W6 W5 W4 W3 W2 W1

W0

LSB

Error

0 MR In3 In2 In1 Out3 Out2 Out1

0 Net MT

Data

Ok

Data

Integrity

TxPDO 2
(T_DW2)

Not in use


Here T_DW1 provides the weight information (32-bit signed integer or 32-bit signed floating point, see chapter
8.6.2; WXX = Weight bit XX; W31 = MSD; W0 = LSD ) and T_DW2 provides the status as indicated below.

Bit no.

TxPDO 2 (T_DW2)

Description

31...29 Error

000

No error

001 ADC

out

010 ADC

over

011 ADC

under

100 System

error

101

FT-11 not in weighing mode

28

Not used

Always 0

27

Multi range (MR)

Single range: always 0

Multi range: 0 = 1. Range / 1 = 2. Range

26

Digital input 3

25

Digital input 2

24

Digital input 1

Status input bit (1 = active)

23

Digital output 3

22

Digital output 2

21

Digital output 1

Status output bit (1 = active)

20

Not used

Always 0

19

Net

0: Gros mode; 1: Net mode

18

Motion

0: No motion; 1: Motion

17

Data Ok

0: Check the error bits; 1: Scale is operating properly

16

Data update in progress

0: PLC should not use the data; 1: use the data
(if 0, the instrument was updating the PLC interface shared memory while data
was read. The PLC interface should ignore this data and rescan.)

15...0

Not used

Always 0

Bit no.

T_DW2

Description during the calibration

0000 0001

Calibration Timeout – Restart calibration

0000 0010

ADC Error – Re-energize the instrument
– If seen again, change board.

0000 0011

Instrument cannot be calibrated – Check load cell cable
– Re-energize the instrument

0010 0010

Instrument cannot be calibrated – Load cell signal is too low or too
high

0010 0011

Calibration Error
– Calibration test weight is too small, increase calibration weight
– Check load cell connections

0010 0101

Scale unstable – Wait until scale becomes stable
– Check grounding wiring

15 … 8

Calibration error

0010 0110

Calibration Jumper is not installed- Check Calibration jumper

0000 0001

Ready for calibration

0000 0011

Zero calibration in process

0000 0100

Span calibration in process

7...0 Calibration

status

0000 1001

Calibration error (see bit 15...8)