FT-11(D) Technical Manual, Rev. 1.35 November 2010

Page 32 of 52

9.

Other I/O Options

9.1. Analogue Output Option (for FT-11 only)

FT-11 can be equipped with 4 – 20 mA or 0 – 10 V analogue output option. The value of the analogue output
changes in linear relationship to the displayed value. But if there is no numerical value on the display, the value
of the analogue output will be given as stated below:

Condition (On Display)

0 – 10 V output

4-20 mA output

Alarm output

The weight value is higher than the weighing range (Over)

13 V

24 mA

High

The weight value is lower than the zero range (Under)

-1.4 V

0 mA

High

Error [Err XX]

13 V

24 mA

High

ADC is out of operating range [Adc Out]

13 V

24 mA

High

As shown in the table above the linear range of the analogue output is limited between -1.4 V and 13 V
respective 0 mA and 24 mA.

9.1.1. Electrical

Connections

There is an optically isolated input on the analogue card for zeroing the scale via an analogue output option
connector. The connections of the analogue output connector is given in the table below.

Pin no. for stainless steel housing

Description

Pin no. for Desk and panel types

(DB25 Female)

J10 connector

(Option 1)

JR1 / JR2 connector

(Option 2 or 3)

+ Zeroing input (+24V)

8

6

13

- Zeroing input ( 0V)

9

5

12

Alarm output

18

8

7

GND (Current / Voltage)

19 & 22

4 & 9

6 & 11

Iout (Current)

23

3

10

Vout (Voltage)

24

2

9

Shield D25

body 1 1

9.1.2. Setup

The analogue output is being adjusted digitally from the parameter group [4--].

[4--] Analogue Output Block

The calibration of the optional analogue output is performed in this sub-block.

[40-] Signal Selection

[400] Signal type*

0 = 4 – 20 mA

1 = 0 – 10 V DC

*activated after saving the settings

[41-] Zero Adjustment (Signal level at load = 0 kg)

[410]

Coarse Zero Adjustment

The coarse zero adjustment is being performed by pressing the <Zero> key continuously to increase the
analogue signal level or the <Clear> key to decrease the analogue signal level. The true load will be
ignored while adjusting.

[411]

Fine Zero Adjustment

The fine zero adjustment is being performed by pressing the <Zero> key repeatedly to increase the
analogue signal level or the <Clear> key to decrease the analogue signal level. The true load will be
ignored while adjusting.

[42-] Span Adjustment (Signal level at maximum load)

[420]

Coarse Span Adjustment

The coarse span adjustment is being performed by pressing the <Zero> key continuously to increase the
analogue signal level or the <Clear> key to decrease the analogue signal level. The full capacity value is
taken as the reference without loading any weight.

[421]

Fine Span Adjustment

The fine span adjustment is being performed by pressing the <Zero> key repeatedly to increase the
analogue signal level or the <Clear> key to decrease the analogue signal level. The full capacity value is
taken as the reference without loading any weight.

Advertising