2 application specific settings – KERN & SOHN KET-TM User Manual

Page 116

Advertising
background image

116

KEN-TM / KET_TM-BA-e-1420

12.2 Application specific settings

Tap icon

in the bottom left corner of the window or call it up from the menu

(For details see chap.8.2).

The application specific settings will be displayed.

Tap button.

Function deactivated

Function activated

Explication:

Ask for multiplier

Enable query, by how many multiples the amount of the recipe

has to be created (such as double the amount = factor 2). The

target weights of the components to be weighed will then be

adjusted accordingly.

Ask for numer of cycles

Query about how often the recipe has to be processed.

Confirm batching

ingredients manually

Each individual component must be imported manually by

pressing the

PRINT key. Tare every time you add a component

by pressing the

TARE key.

Automatic tare

Automatic import of individual components after stability has

been established.
Automatic taring on start of recipe and every time a component

is added.

Ingredient control

When “

Check Component is enabled the editing window

<Check Component> will appear prior to adding the

component.

The editing window is used to enter the component’s code with

the help of a bar scanner.

Portion weighing

To reach the target weight, enable “Dosing of Components in

Random Dosages“ mode.

Record printout

Automatic log printout after recipe completion

Advertising