18 Troubleshooting guide / error messages

Possible causes of errors:

In case of an error in the program process, briefly turn off the balance and disconnect

from power supply. The weighing process must then be restarted from the beginning.
Fault

Possible cause

The displayed weight does

not glow.

The display unit is not switched on.

Mains power supply interrupted (mains cable

defective).

Power supply interrupted.

The displayed weight is

permanently changing

Draught/air movement

Table/floor vibrations

Weighing pan has contact with other objects.

Electromagnetic fields / static charging (choose

different location/switch off interfering device if

possible)

The weighing result is

obviously incorrect

The display of the balance is not at zero

Adjustment is no longer correct.

Great fluctuations in temperature.

Warm-up time was ignored.

Electromagnetic fields / static charging (choose

different location/switch off interfering device if

possible)

No data transfer between

printer and balance.

Communication settings are wrong.

The menu setting cannot be

changed.

Menu item is locked for models with type approval

certificate.

Error messages:

Err2

Value outside zero setting range

Err3

Value outside tare range

Err8

Taring or resetting to zero when time is exceeded during

procedure

Er4 FuL2

Weighing range exceeded

NULL

Error AD transformer

FULL

Weighing range exceeded

HI

Display range exceeded

LH

Initial weight error