Delete tare
1. Unload weighing plate and press
.
The (
Net) indicator turns off, the zero display shows.
2. When the weighing platform is loaded, you can only press
if the zero setting
range is kept within 2% max (>2% max will generate error message “Err 2“).
The (
Net) indicator turns off, the zero display shows.
3. Press function key
<Turn off tare> when weighing platform is loaded (<2%
max) or unloaded (For allocation of function key see chap.7.4.2. The (
Net)
indicator turns off, the zero display shows.
If there is a function key allocated to
<Reset tare>
, the most recently used
tare value will be called up. (Allocation of function key, see chap. 7.4.2).
6.10 Connection of peripheral devices
Before connecting or disconnecting of additional devices (printer, PC) to the data
interface, always disconnect the balance from the power supply.
With your balance, only use accessories and peripheral devices by KERN, as they are
ideally tuned to your balance.