4 configure display unit – KERN & SOHN KET-TM User Manual

Page 196

Advertising
background image

196

KEN-TM / KET_TM-BA-e-1420

21.4 Configure display unit

Before configuring a display unit make sure you know the required configuration data,

see chap. 21.2. The weighing system is equipped with a service switch protected

service level for entering configuration data.

Call up service level:

Switch on device with

To call up the service level you have to operate the service switch.

Use a suitable instrument for pushing the service switch

(such as thin screwdriver, Ø approx. 3 mm).

Fig.: Position

service switch

Attention:
To operate the service switch of calibrated weighing systems you will first have to

destroy the seal.

After destruction of the seal the weighing system must be re-verified by an

authorised agency and a new verification wire/seal mark fitted before it can be

reused for applications subject to verification.

Call up menu with

and the complete menu (See chap. 2.3.1) incl. service level

will be displayed. The two service blocks of the service level are described below:

<

General >

<

Manufacturer parameters >