KEN-TM / KET_TM-BA-e-1420

16.3 Detailed information about interface logs

16.3.1 Zeroing

Befehl:

Z CR LF

Possible responses:


Z

˽

A CR LF

Z

˽

D CR LF

Command accepted; executing command
Execution of previously started command completed

Z

˽

A CR LF

Z

˽

^ CR LF

Command accepted; executing command
Command valid but zero setting range was exceeded

Z

˽

A CR LF

Z

˽

E CR LF

Command accepted; executing command
Time limit for stabilisation of weighing scale display was exceeded

Z

˽

CR LF

Command valid but can currently not be executed

16.3.2 Taring

Command:

T CR LF

Possible responses:


T

˽

A CR LF

T

˽

D CR LF

Command accepted; executing command
Execution of previously started command completed

T

˽

A CR LF

T

˽

^ CR LF

Command accepted; executing command
Command valid but tare range was exceeded

T

˽

A CR LF

T

˽

E CR LF

Command accepted; executing command
Time limit for stabilisation of weighing scale display was exceeded

T

˽

I CR LF

Command valid but can currently not be executed