21 installation display unit / platform, 1 technical data, 2 weighing system design – KERN & SOHN KET-TM User Manual

Page 193

Advertising
background image

KEN-TM / KET_TM-BA-e-1420

193

21 Installation display unit / platform

Installation / configuration of a weighing system must be carried out by a well

acquainted specialist with the workings of weighing balances.
To configure a weighing system you have to call up the service level in the

menu. Here you can change all parameters. For this reason direct access to the

service switch is locked.
For disabling access lock see chap. 21.4.

21.1 Technical data

KET-TM

KEN-TM

Supply voltage platform

5 V

Max. signal voltage

19.5 mV

Load cell resistance

min. 80 Ω

max. 1200 Ω

Max number of platforms

2

4

21.2 Weighing system design

The display unit is suitable for connection to any analogue load cell in compliance with

the required specifications.

The following data must be established before selecting a load cell:

Weighing balance capacity

This usually corresponds to the heaviest load to be weighed.

Preload

This corresponds to the total weight of all parts that are to be placed on the

weighing cell such as upper part of platform, weighing pan etc.

Zeroing range

Smallest desired display division

Verifiable, if required

The addition of weighing scales capacity, preload and the total zero setting range give

the required capacity for the weighing cell.

To avoid overloading of the weighing cell, include an additional safety margin.

Advertising
This manual is related to the following products: