KEN-TM / KET_TM-BA-e-1420

21.4.2 Adjusting a weighing system

Tap

< Adjustment >

The menu will be displayed

Determine initial

mass

Adjust zero point

Calibration

How to carry out

adjustments

Initial mass

A/D transformer at zero

Calibration factor Adjusting co-efficient

Adjustment

weight

For mass of required

adjustment weight see list of

parameters “Configuration”.

How to carry out adjustment:

Prepare the required adjustment weight. The weight to be used depends on the

capacity of the scale. Carry out adjustment as near as possible to the scale’s

maximum weight. Info about test weights can be found on the Internet at:

http://www.kern-sohn.com.

Observe stable environmental conditions. Stabilisation requires a certain warm-

up time.

When using several platforms follow the sequence of steps below for each

platform.