KEN-TM / KET_TM-BA-e-1420

16.3.9 Send weighing value immediately in current weighing unit

Command:

SUI CR LF

Possible responses:


SUI

˽

I CR LF

Command valid but can currently not be executed

1 2 3

4

5

6

7-15

16

17-19

20

21

S U I Stability

icon

˽

Signs

Weight

˽

Unit

CR

LF

Example:

S U I

CR LF

For sent command see chap. 16.1

S U I ?

˽

-

˽˽˽

5 8 . 2 3 7

˽

k g

˽

CR LF

Command executed.

Weighing value will issued in current weighing unit

16.3.10

Start continuous output in standard weighing unit

Command:

C1 CR LF

Possible responses:


C1

˽

I CR LF

Command valid but can currently not be executed

C1

˽

A CR LF

Command accepted; executing command
Weighing value will be issued in default weighing unit

1 2 3

4

5

6

7-15

16

17-19

20

21

S I

˽

Stability

icon

˽

Signs

Weight

˽

Unit

CR

LF

16.3.11

Stop continuous output in standard weighing unit

Command:

C1 CR LF

Possible responses:


C0

˽

I CR LF

Command valid but can currently not be executed

C0

˽

A CR LF

Command accepted; executing command