KEN-TM / KET_TM-BA-e-1420

7.1.3 Auto Zero

The automatic zero point adapter (auto zero) provides continuous

correction for minor weight variations such as contamination on the

weighing platform.

In the event that small quantities are removed or added to the material to

be weighed, incorrect weighing results can be displayed due to the

“stability compensation“. (e.g. slow flow of liquids from a container placed

on the balance, evaporating processes).

When apportioning involves small variations of weight, it is advisable to

switch off this function.

Available settings:

- Auto zero disabled

- Auto zero enabled

7.1.4 LO threshold

Entering threshold for automatic operation

The next measurement will not be saved until the weighing display falls

below the net value set for the LO threshold.

Tap

<LO threshold> and the numeric

input window will appear.
Enter value and take over by

.

7.1.5 Last digit display accuracy

Lower display accuracy results in faster display.

Tap

< Last digit >

The menu will appear.

Always

All digits on

Never

Last digit off

When stable Last digit on for

stable weighing value

Advertising