3 retrieve tare value, 4 set tare value – KERN & SOHN KET-TM User Manual

Page 169

Advertising
background image

KEN-TM / KET_TM-BA-e-1420

169

16.3.3 Retrieve tare value

Command:

OT CR LF

Answer:

1

2

3

4 - 12

13

14 - 16

17

18

19

O

T

˽

Tare

˽

Unit

˽

CR

LF

Tare value: - 9 characters with right-justified alignment

Unit:

- 3 characters with left-justified alignment

Tare values are always issued in adjusting unit

16.3.4 Set tare value

Command:

UT

˽

TARA CR LF, (TARA = tare value)

Possible responses:

UT

˽

OK CR LF

Command executed

UT

˽

I CR LF

Command valid but can currently not be executed

ES CR LF

Invalid entry

Set tare value:

Mark decimal places with a dot

Without weighing unit

16.3.5 Send stable weighing value in standard weighing unit

Command:

S CR LF

Possible responses:

S

˽

A CR LF

S

˽

E CR LF

Command accepted; executing command

Time limit for stabilisation of weighing scale display was exceeded

S

˽

I CR LF

Command valid but can currently not be executed

1

2-3

4

5

6

7-15

16

17-19

20

21

S

˽

Stability

icon*

˽

Signs

Weight

˽

Unit

CR

LF

Advertising