16 retrieve value for “lower threshold, 17 retrieve value for “upper threshold, 18 press simulation “print key – KERN & SOHN KET-TM User Manual

Page 174: 19 send all implemented messages

Advertising
background image

174

KEN-TM / KET_TM-BA-e-1420

16.3.16

Retrieve value for “Lower Threshold“

Command:

ODH CR LF

Possible responses:


DH

˽

MASA CR LF

Command executed

1

2

3

4-12

13

14-16

17

18

19

D

H

˽

Weight

˽

Unit

˽

CR

LF

Weight

- 9 characters with right-justified alignment

Unit

- 3 characters with left-justified alignment

16.3.17

Retrieve value for “Upper Threshold“

Command:

OUH CR LF

Possible responses:


UH

˽

MASA CR LF

Command executed

1

2

3

4-12

13

14-16

17

18

19

D

H

˽

Weight

˽

Unit

˽

CR

LF

Weight

- 9 characters with right-justified alignment

Unit

- 3 characters with left-justified alignment

16.3.18

Press simulation “PRINT key“

Command:

SS CR LF

Using the command

S S CR LF will save the weighing activity automatically to the

database and a printout will follow provided an optional printer is connected.

16.3.19

Send all implemented messages

Command: PC CR LF

Answer:

PC A

"Z,T,S,SI,SU,SUI,C1,C0,CU1,CU0,DH,ODH,UH,OUH,OT,UT,SIA,SS, PC"