3 solid state temperature and humidity probe, Specification, Sensor specifications – Casella CEL Indoor air quality monitor User Manual

Page 31

Advertising
background image

12.3 Solid State Temperature

and Humidity Probe

This sensor combines two transducers into a
single probe allowing for the measurement of both
temperature and relative humidity to be made
simply and quickly.

The relative humidity transducer is based on a
Capacitive element using “Thin Film” technology,
while the temperature transducer uses a Ceramic
based “Platinum Resistance Wire” technology.

The relative humidity readings are temperature
compensated to reduce the error due to changes
in ambient temperature.

Specification

Relative Humidity

Type

Capacitive

Operating Range

10% to 98% RH

Calibration Accuracy

±2% RH (inc. hysteresis)

Response Time (T

90

)

2.5 S

Calibration

Check once per year

Calibration Hysteresis

±5%

Hysteresis at 20

o

C

±5% RH

Long Term Stability

Better than 15 over one year

Calibration Check Points

35%, 80% & 10% RH

Overall Total Accuracy

±3% RH

Linear Voltage Output

0 to 1 V DC for 0 to 100% RH

Temperature Transducer

Type
Operating Range
Calibration Accuracy
Response Time (T

90

)

Calibration
Calibration Hysteresis
Linear Voltage Output

Sensor Specifications