Sensor Specifications

Operating Time

>10 hours with NiCd on full charge

Weight

Approx. 1 kg

Analogue Output

0 - 2.5 V

Infra-Red Light Source

980 nm wavelength

12.7 CO Gas Detector

The CO gas detector has been designed to provide continuous pollution
monitoring. The sensor uses electrolytic fuel cells which have a typical life
expectancy of 2 to 3 years, depending on the working conditions (temperature,
humidity, etc.).

The sensor is designed to detect
the specific gas with a minimum
concentration of 0.1 ppm. For
applications where the gas is
anticipated in any appreciable
quantity, this level of detection is
sufficient.

For ambient gas detection at a
distance from the source, a gas
analyser may be required which
has the capability to measure
down to levels of 1 ppb.

The standard detector is housed
within a Polycarbonate Housing
and can be mounted on the hub
support arms with spring clips.
The units must be installed in
non-classified areas.

Sensor

02065

Slide-in
labels

Connection
cable

Clips

Gas Detector

CTX300

Location

Gas

CO

Specification

Range

0 - 300 ppm

Accuracy

±1.5%

Response Time T(50)

15 seconds

Operating Humidity Range

10 - 95% RH

Operating Temperature Range

-20 to +50

o

C

Supply Voltage

15 - 32 V

Atmospheric Pressure

±10%

Life Span

24 to 36 months

Housing

Sealed to IP66

When Ozone measurement is required together with CO, a
special connector box will be required to connect the two
sensors into one

channel.