12.5 CO

2

Monitor

This monitor provides a robust and accurate
method for obtaining CO

2

measurement in site

survey applications.

The sensor is based on an “Infra-red Absorption”
technique. The unit incorporates a single beam
emitter, where the single beam is split into two
using solid state circuitry. This provides a dual
wavelength infra-red sensing chamber with no
moving parts, which results in a low drift, high
accuracy monitor with fast response and low
power consumption.

The unit incorporates a diaphragm pump to draw
the sample air into the measurement chamber,
thus reducing the response time as compared to a
diffusion type of unit. A display on the front of the
instrument provides the user with real time
readings of the CO

2

concentration.

Sensor Specifications

Specification

Operating Range

Standard = 0 to 3000 ppm (0.3%)
Options: 1%, 3%, 10%, 30%, 100%

Power Required

24 V DC (range 18 to 30 V DC) at 6 W
(250 mA)

Response Time (T

90

)

10 S

Operating Temperature

0

o

C to 45

o

C

Warm Up Time

On initial power up, 30 S

Humidity Effects

Unaffected by humidity (non-condensing)

Sensor Protection

Internal particle filter

Operation

An integral pump is fitted that draws air
through the sensing head

Air Flow Control

The pump is flow controlled to ±5% of a
nominal 0.5 l/min (with max. back
pressure of 10 cm H

2

O

Accuracy

±2% of full scale (calibration gas
accuracy not included)

Stability

±2% of full scale over 12 months

Hysteresis

Zero ±0.3% - Span ±1.5%

Zero Drift at Constant Temp. ±2% full scale per 12 months
Zero Drift at Ambient Temp.

±0.03% full scale per

o

C

Accuracy over calibrated
period as a % of full scale

60 ppm (cal. per year)
30 ppm (cal. 6 monthly)
15 ppm (cal. 3 monthly)

Page 33 of 42