Gas detector, Sensor specifications, Specification – Casella CEL Indoor air quality monitor User Manual

Page 37

Advertising
background image

Sensor Specifications

12.8 O

2

Gas Detector

The O

2

gas detector has been designed to provide continuous pollution

monitoring. The sensor uses electrolytic fuel cells which have a typical life
expectancy of approximately 2 years, depending on the working conditions
(temperature, humidity, etc.).

The sensor is designed to detect
the specific gas with a minimum
concentration of 0.1 ppm. For
applications where the gas is
anticipated in any appreciable
quantity, this level of detection is
sufficient.

For ambient gas detection at a
distance from the source, a gas
analyser may be required which
has the capability to measure
down to levels of 1 ppb.

The standard detector is housed
within a Polycarbonate Housing
and can be mounted on the hub
support arms with spring clips.
The units must be installed in
non-classified areas.

Sensor

02066

Slide-in
labels

Connection
cable

Clips

Gas Detector

CTX300

Location

Gas

O

2

Specification

Range

0 - 30%

Accuracy

±1.5%

Response Time T(50)

10 seconds

Operating Humidity Range

10 - 95% RH

Operating Temperature Range

-20 to +50

o

C

Supply Voltage

15 - 32 V

Atmospheric Pressure

±10%

Life Span

20 to 28 months

Housing

Sealed to IP66

When Ozone measurement is required together with CO, a
special connector box will be required to connect the two
sensors into one

channel.

Page 37 of 42

MICROTHERM indoor air quality &
WinIaq Application Software - User Manual