Chapter 8 : Model 2612 Strain Gauge/RTD Module

Command

(Adrs),(Data)

Response

none

Notes

This function is used by the middleware to store
calibration data for the module. Refer to the
distribution media for details.

Due to the lengthy time required to write a byte to
the EEPROM, this action should be invoked only
once per command packet.

8.8 Specifications

Parameter

Function

Adrs

Memory address that is to be written to.

Data

Byte value that is to be stored.

Table 25: Model 2612 Specifications

Analog Inputs

Resolution

24 bits

Input channels

4

Conversion rate

6.875 to 3500 samples/second for each channel.

Input range

±1/2 of reference.

Input impedance

110 kOhm, minimum, measured from either sensor input to reference voltage.

Power Outputs

Output channels

4

Output power

1.25V, 2V, 3V or 5V, ±10%@ 1 to 25 mA for each channel.

Mating Connectors
(*) = included with module

IOM port (qty 1)

RJ-45 plug, AMP 554169 or equivalent

Analog inputs (qty 4)

RJ-45 plug, AMP 554169 or equivalent

Analog inputs (qty 4)

(*)Pluggable TB, 8-pin, RIA 31166110 or equivalent

Temperature

Operating range

0 to 70°C

Power

Input power

+24V ±5% @ 60mA.

Dimensions

Outer dimensions

5.0H x 5.0W x 1.8D inches, including DIN mounting frame. Depth is specified with
respect to surface of DIN rail support panel; additional clearance required for field wiring.