Centroid, Elliptical pass/fail, Orientation – SIGMA LBA-708 User Manual

Page 100: Gauss pass/fail, Top hat pass/fail

Advertising
background image

3.3.2.2 Centroid

The Centroid Pass/Fail item allows you to define a circle that must contain the centroid of the

beam energy. To implement this test you must define the location of the center of a circle in
terms of its X and Y coordinate in the beam display window, and the Radius of the circle that

must contain it.
Both the X and Y Centroid location results will change color from

GREEN

to

RED

should the

centroid fall outside this circle.

3.3.3 Elliptical

Pass/Fail

The following results may be set in the Pass/Fail Elliptical… dialog below the main menu

Pass/Fail item.

Figure 43

3.3.3.1 Orientation

The Orientation Pass/Fail item applies to the major axis inclination of an Elliptical beam. The

units for these two edit controls are always in degrees. The Major axis can tilt through an angle
in the range of +90 to -90 degrees. You can set the desired tilt Angle, and a +/- Range around

this angle, that will meet your testing criteria. The Rotation result will change color from

GREEN

to

RED

should the angle fall outside of these set limits.

3.3.4 Gauss

Pass/Fail

See the Pass/Fail Gauss… dialog below the main menu Pass/Fail item for a list of these items.

3.3.5

Top Hat Pass/Fail

See the Pass/Fail Top Hat… dialog below the main menu Pass/Fail item for a list of these items.

Operator’s Manual

LBA-PC

100

Advertising