Whole beam fit equations, X/y or major/minor line fit equations, 14 whole beam fit equations – SIGMA LBA-708 User Manual

Page 136: 15 x/y or major/minor line fit equations, Ej j

Advertising
background image

(

)

∑∑

=

x

y

xy

xy

S

Z

A

2

min

Where:

Z

xy

= Amplitude of the pixel data at (x,y).

S

xy

= Amplitude of fitted surface at (x,y).

6.14 Whole Beam fit equations

The bivariate normal equation is used to fit data in two locked directions, X and Y. The Whole Beam

selection assumes the beam is round or elongated parallel to the horizontal or vertical axis. The
definition of the bivariate normal equation and the displayed results are as follows:



+





 −

=

2

2

2

/

2

/

2

y

w

y

y

x

w

x

x

e

J

J

o

Where:

J

=

Amplitude at the point (

x,y

).

Jo

* =

Amplitude at the Gaussian center.

x

=

x

location of pixel.

x

* =

x

location of the Gaussian center.

wx

* =

Horizontal width at 1/e² of energy.

y

=

y

location of pixel.

y

* =

y

location of the Gaussian center.

wy

* =

Vertical width at 1/e² of energy.

Parameters marked with an asterisk (*) are the variables fitted.

6.15 X/Y or Major/Minor line fit equations

The univariate normal equation is used to fit data in one direction. The definition of the equation and

the displayed results are shown below:
for the X or Major axis

2

2

/

2





=

M

w

M

M

M

e

J

J

Where:

J

=

Amplitude at the point

M

.

J

M

*

=

Amplitude at the Gaussian center.

M

=

Location of pixel.

M

=

location of the Gaussian center.

w

M

* =

Width at 1/e² of energy.

Operator’s Manual