Operator’s Manual

LBA-PC

222

BMP

B

.BMP logging enable.

Key Type Value

Description

DataLogging

B

Data logging enable.

DataFileName

S

Name of data log file.
If no extension (i.e., no period) then

.LB3/4/5/7 is appended.
Maximum 256 characters.
If this key is set to “FRM” or “RDD” (case

ignored) (or was previously set to FRM or

RDD in the dialog) then data frames will be

logged to the GPIB bus. Each time a new

frame is captured the LBA-PC will

automatically send a response as if you had
sent “:FRM?” or “RDD?” respectively. You

must address the LBA-PC to talk and be

prepared to accept the transmitted data.

The LBA-PC will wait until you have read the

data before capturing another frame.

Note:

FRM and RDD are ignored if the LBA-PC is

not in remote.

ResultsLogging

B

Results logging enable.

ResultsFileName

S

Name of results log file.
If no extension (i.e., no period) then .RLG is

appended.
Maximum 256 characters.
If this key is set to “RDR” (case ignored) (or

was previously set to RDR in the dialog) the

results will be logged to the GPIB bus. Each

time results are computed for a new frame

the LBA-PC will automatically send a

response as if you had sent “:RDR?
Values=1”. The LBA-PC will wait until you

have read the data before capturing another

frame.

Note, in this case the results format

is ignored. Note: RDR is ignored if the LBA-

PC is not in remote.

ResultsFormat

L

Results log format.
0 = math
1 = spreadsheet

ExportLogging

B

Export logging enable.

ExportFileName

S

Name of export log file. Any file extension is

ignored and an appropriate extension is used

depending on the type of export file that is

being logged.