SIGMA LBA-708 User Manual

Page 91

Advertising
background image

Figure 37

The Centroid and Peak Histogram windows now have horizontal and vertical plot scaling in units

of pixel, with histogram bins the size of a single pixel. Note also that we have the same plot

orientation as the LBA-PC beam window. In the Figure above we can see that most of the

centroids are falling 232

pixels from the left and 247 pixels from the bottom (assuming that your

origin is set to Window LL). In order to get this into real world units we would simply multiply
the pixel units by the pixel scale of the camera. In the example above the pixel scale, read form

the camera dialog was 13 microns per pixel. Thus the centoid would become roughly:

)

(

58

.

3024

13

)

(

66

.

232

m

pixel

m

pixel

µ

µ

=





From the left (X)

)

(

4

.

3221

13

)

(

80

.

247

m

pixel

m

pixel

µ

µ

=





From the bottom (Y)

The ellipses in the histogram windows have X and Y diameters two standard deviations wide and

tall respectively.

Operator’s Manual

LBA-PC

Doc. No. 10654-001, Rev 4.10

91