Ieee 488.2 common commands – SIGMA LBA-708 User Manual

Page 212

Advertising
background image

Operator’s Manual

LBA-PC

212

whenever a pass/fail alarm occurs while

running.

A.2 IEEE 488.2 Common Commands

The following IEEE 488.2 common and related commands are supported by the LBA-PC.

Command Meaning

Usage

*IDN?

Identification

LBA-PC returns the string “Spiricon, LBA-

PC, ssss, v.vv” where ssss is the serial
number of the board and v.vv is the

software version number.

*RST

Reset

Causes the LBA-PC to do the following:
Clear ELR, ESR, STB
Stop running
Restore last configuration saved or

restored

:CLR Clear

Queues

and Status

Registers

Clear output queue (MAV) and error

message queue (EMAV). Clear ELR, ESR,

STB.

*CLS Clear

Status

Registers

Clear ELR, ESR, STB.

:ELE 300PC

Event

Status Enable

Set mask to enable event status

notification of corresponding bits set in

ELR.

:ELE?

Return ELE contents.

:ELR? 300PC

Event

Status
Register

Return and clear ELR contents:
bit 7 - Range Error flag, set when host
attempts to set an input to a value that

exceeds permissible range.
bit 6 - Checksum Error flag, set when

locally computed checksum does not

match transmitted value.
bit 5 - unused
bit 4 - unused
bit 3 - unused
bit 2 - New Frame Available flag, set each

time the LBA-PC has acquired a new

frame of data.
bit 1 - Results Available flag, set each
time the LBA-PC completes a set of

computations for a new frame of data.
bit 0 - Results Pass/Fail flag, set

Advertising
This manual is related to the following products: