Pass/fail results – SIGMA LBA-708 User Manual

Page 168

Advertising
background image

1 Standard

Deviation

2 Minimum
3 Maximum

All results values will be loaded into the array all the time. Results not enabled in the LBA-PC will

be zero. Note there are three sets of Gauss Fit and Top Hat results. Either the whole beam

section or the major and minor sections will contain valid results depending on how the LBA-PC is

configured.

9.3.1.23 Pass/Fail

Results

The PassFailFlag and PassFail properties provide all of the LBA-PC pass/fail test results.
The PassFailFlag is a combination of all the individual pass/fail results. The value of PassFailFlag

is defined as follows:

Fail

Some test failed

No test

No individual test enabled

Pass

All tests passed


The PassFail property provides individual pass/fail results in a one-dimensional array of integers

packaged as a Variant. The pass/fail results are loaded into the array in the following order:

1. Quantitative

• There is no test for Peak Location X or Y. The value of both of these flags is always zero.

• There is no test for Radius so the flag is always zero.

2. Elliptical
3. Gauss Fit – whole beam
4. Gauss Fit – major axis
5. Gauss Fit – minor axis
6. Top Hat – whole beam
7. Top Hat – major axis
8. Top Hat – minor axis
9. Divergence

This is the same order as listed in sections 3.1.15 through 3.1.19 above under individual results
properties.
Each ‘flag’ will contain one of the following values:

-1 Fail

Operator’s Manual

LBA-PC

168

0 No

Test