Key Type Value

Description

FrameNumber I Frame

number

-1 = gain frame
0 = reference frame
1 to n. ‘n’ is the size of the frame buffer.
default = current frame

Row

I

1 to h. ‘h’ is the frame height.
default = current cursor location


Returns:

RCR FrameNumber=f;Row=r;#dn..n(DAW)( DAW)…(DAW)(DAW)


Where:

f = frame

number

r

=

which row is being returned

# = pound

symbol

d

=

number of digits to follow in n..n

n..n

=

number of data words (i.e. row length)

(DAW) =

data word is two 8-bit data bytes, low byte followed by high byte

Each data word is a two’s complement fixed point value in one of the following formats:

siiiiiii ifffffff

LBA-300PC

-256 to 255.9921875

siiiiiii iiifffff

LBA-400PC

-1024 to 1023.96875

siiiiiii iiiiifff

LBA-500PC

-4096 to 4095.875

siiiiiii ifffffff

LBA-708PC

-256 to 255.9921875

siiiiiii iiifffff

LBA-710PC

-1024 to 1023.96875

siiiiiii iiiiifff

LBA-712PC

-4096 to 4095.875

siiiiiii iiiiiiif

LBA-714PC

-16384 to 16383.5

Where:

s = sign

bit

I =

integer

f =

fraction

Operator’s Manual