Log - logging – SIGMA LBA-708 User Manual

Page 197

Advertising
background image

Centroid

X,

Centroid

Y,

Width

X,

Width

Y,

Diameter(^END)

LBA-PC sends

RDR

,%,,,um,um,um,um,um,um,um(^END)

10.6.2.6 LOG - Logging

You can configure the LBA-PC to automatically send a FRM or RDD and RDR response each time

new data is acquired. Remote logging is the fastest method available to transfer new frames

from the LBA-PC.
Use the File | Logging… dialog box or the LOG configuration command to specify logging new

data or results to the host. If the data log file name is “FRM” or “RDD” then each time a new

frame of data is acquired the LBA-PC will automatically send a response as if you had sent

“:FRM?” or “:RDD?” respectively. If the results log file name is “RDR” then each time a new

frame of data is acquired the LBA-PC will automatically send a response as if you had sent

“:RDR?”.
The LBA-PC will not capture another frame until the host reads the current logging response.
This example describes how to enable data file logging to the host controller.

Host sends
:LOG
DataLogging=1;
DataFileName=FRM(^END)

Host sends
:RUN

LBA-PC sends
FRM

FrameNumber=1;
#516384(DAB)…(DAB)

(^END)


LBA-PC sends

Operator’s Manual

LBA-PC

Doc. No. 10654-001, Rev 4.10

197

FRM
FrameNumber=2;

Advertising