The following example describes how to download the current result values displayed in the

results window:
Host sends

:RDR?

(^END)

LBA-PC sends

RDR

3298655,
86.96,
1.776e+02,
2.288e+01,
1.846e+03,
1.950e+03,
1.696e+03,
1.519e+03,
3.842e+03,
3.781e+03,
3.812e+03(^END)

The following example describes how to download the current labels and units displayed in the
results window.

Host sends

:RDR?

Labels=1;
Values=0;
Units=1(^END)

LBA-PC sends

RDR

Total,

% Above Clip,Peak,

Min,

Peak Loc X,