I

Integer. ASCII numeric value in integer format.

B

Boolean. ASCII numeric integer value.
0 = false. 1 = true.

F

Fixed/Floating. ASCII numeric value in fixed or floating point

notation.

S

String. Series of ASCII characters. Note that the backslash,

“\”, has special meaning in strings known as an escape

sequence. To specify a single backslash, you must send two.

For example: “c:\\spiricon\\lba300pc\\lbapc.cfg” is interpreted

as “c:\spiricon\lba300pc\lbapc.cfg”.

Although other escape

sequences are recognized there is no reason ever to use them

with LBA-PC.

T

Time. ASCII numeric integer values in the form
[[HHH:]MM:]SS.

D

Date. ASCII numeric integer values in the form MM/DD/YY.


You can obtain the current settings associated with any configuration command by sending the

command followed by the query character, “?”. Any parameters that are part of a query command is

a command error, unless they are specifically allowed in the command description.
Configuration query commands return all of the keys for the specified configuration in the following

format:
CCC

key=value;key=value;…;key=value;key=value;(^END)

Other query commands respond in ways that are specific to a particular command. See the individual

command listings in Appendix A for details.
The following example illustrates the process for setting the Crosshair Location Mode to Peak.

Host sends
:DIS

Crosshair=3(^END)

The following example illustrates the process for querying the current Aperture configuration.

Host sends
:APT?(^END)

LBA-PC sends

APT

DrawShape=1;
CenterXLoc=1.600e+01;
CenterYLoc=1.500e+01;

Operator’s Manual

LBA-PC

184

Major=8.000e+00;