Frm - upload/download a data frame – SIGMA LBA-708 User Manual

Page 250

Advertising
background image

Message Generated

By

%s detected but cannot be initialized. LBA-PC set

to Off-Line mode.

device driver

Unable to load LCA program file (%s).

device driver

Press OK to stop Auto Calibration.

ultracal

A.5.4.6 FRM - upload/download a data frame

Only one frame at a time can be uploaded or downloaded. :FRM will upload a frame of data
from the controller to the LBA-PC, while :FRM? will download a frame of data from the LBA-PC to

the controller. Data is uploaded to or downloaded from the LBA-PC’s frame buffer.

:FRM? <configuration> (LBA300PC

→ CONTROLLER)

Key Type Value

Description

FrameNumber I frame

number

-1 = gain frame
0 = reference frame
1 to n. ‘n’ is the size of the frame buffer.
default = current frame

Returns:

FRM FrameNumber=f;#dn..n(DAB)(DAB)…(DAB)(DAB)

Where:

f = frame

number

# = pound

symbol

d

=

number of digits to follow in n..n

n..n

=

number of data bytes

(DAB) =

data byte - 8-bit data byte


:FRM [FrameNumber=f;][Replace=B;]#dn..n(DAB)(DAB)…(DAB)(DAB)
:FRM <configuration> (CONTROLLER

→ LBA300PC)

(DAB)(DAB)…(DAB)(DAB) must be exactly what was returned by a previous :FRM? command.
Where:

f = frame

number

# = pound

symbol

Operator’s Manual