SIGMA LBA-708 User Manual

Page 207

Advertising
background image

The second category of errors is LBA-PC execution errors. These error messages are normally

displayed in a message box on the screen. The LBA-PC reroutes these error messages to the error

message queue when the queue is enabled.
Example of a category 2 error:
Host sends

:PAN

X=0;
Y=0(^END)


Host sends

*STB?(^END)


LBA-PC sends

STB

#H08(^END)


Host sends

:ERR?(^END)


LBA-PC sends

!!! PAN - ‘X=0’ out of range (112,744)(^END)

The third category of errors are either fatal or result from some local user action. These errors cannot

occur because of remote commands. This third category of errors is always displayed in a message box
on the screen and requires user action before continuing. See the ERR specification for the specific

errors included in each error category.
Example of a category 3 error:
Operator clicks START!
Host sends

:LDC


Host sends

*STB?(^END)


LBA-PC sends

STB

Operator’s Manual

LBA-PC

Doc. No. 10654-001, Rev 4.10

207

#H08(^END)

Advertising