Differential connections for ground-referenced sig, Figure 2-7. differential input connections for gro, Sources -23 – National Instruments AT-MIO-16X User Manual

Page 45: Figure 2-7

Advertising
background image

Chapter 2

Configuration and Installation

© National Instruments Corporation

2-23

AT-MIO-16X User Manual

Differential Connections for Ground-Referenced
Signal Sources

Figure 2-7 shows how to connect a ground-referenced signal source
to an AT-MIO-16X board configured in the DIFF input mode. The
AT-MIO-16X analog input circuitry must be configured for DIFF input
to make these types of connections. Configuration instructions are
included in Chapter 4, Register Map and Descriptions.

Figure 2-7. Differential Input Connections for Ground-Referenced Signals

With this type of connection, the PGIA rejects both the common-mode
noise in the signal and the ground potential difference between the
signal source and the AT-MIO-16X ground, shown as V

cm

in

Figure 2-7.

Ground-
Referenced
Signal
Source

Common-
Mode
Noise,
Ground
Potential,
and so on

Input Multiplexers

AI SENSE

PGIA

V

m

Measured
Voltage

AI GND

-

+

-

+

V

s

V

cm

+

-

-

+

I/O Connector

AT-MIO-16X Board in the DIFF Input Configuration

ACH<0..7>

ACH<8..15>

Gain

Advertising