Chapter 2

Configuration and Installation

© National Instruments Corporation

2-31

AT-MIO-16X User Manual

Digital I/O Signal Connections

The digital lines ADIO<0..3> are connected to digital I/O port A. The
digital lines BDIO<0..3> are connected to digital I/O port B. DIG GND
is the digital ground pin for both digital I/O ports. Ports A and B can be
programmed individually to be inputs or outputs.

The following specifications and ratings apply to the digital I/O lines.

Absolute maximum voltage input rating

5.5 V with respect to
DIG GND

Digital input specifications (referenced to DIG GND):

V

IH

input logic high voltage

2 V minimum

V

IL

input logic low voltage

0.8 V maximum

I

IH

input current load,

logic high input voltage

40

µ

A maximum

I

IL

input current load,

logic low input voltage

–120

µ

A maximum

Digital output specifications (referenced to DIG GND):

V

OH

output logic high voltage

2.4 V minimum

V

OL

output logic low voltage

0.5 V maximum

I

OH

output source current,

logic high

2.6 mA maximum

I

OL

output sink current, logic low

24 mA maximum

With these specifications, each digital output line can drive 11 standard
TTL loads and over 50 LS TTL loads.

Figure 2-12 depicts signal connections for three typical digital I/O
applications.