Light source controls for the main tester, Ac b – Fluke Networks CertiFiber Pro Users Manual User Manual
Page 237

Chapter 8: The Power Meter and Light Source
How to Control the Light Source
215
GPU127.EPS
Figure 94. Light Source Controls for the Main Tester
Use the buttons to control the output port when you use the
main tester as a light source:
Auto: In this mode, the output port transmits both
wavelengths. The signal includes identifiers that tell a
CertiFiber Pro or SimpliFiber Pro power meter when to
measure power at each wavelength. Use this mode only
with a CertiFiber Pro or SimpliFiber Pro power meter.
Off: The output port does not transmit the selected
wavelength.
CW: The output port transmits a continuous signal at one
wavelength. Use this mode if the power meter is not a
CertiFiber Pro or SimpliFiber Pro meter.
The laser icon is red when the output port transmits the
wavelength adjacent to the icon.
When the tester has a CertiFiber Pro Quad module installed, use
this control to select multimode or singlemode wavelengths
A
C
B