Ph 1 – Xylem HandyLab 600 User Manual

Page 25

Advertising
background image

HandyLab 600

pH value

ba77064e01

03/2014

25

Continuing with two-

point calibration

4.

Thoroughly rinse the sensor with deionized water.

5.

Immerse the sensor in the first buffer solution.

6.

For measurements without temperature sensor :
Enter the temperature of the buffer with <><>.

7.

Start the measurement with <ENTER>.
The measured value is checked for stability (stability control).
The [AR] status indicator is displayed. The measured parameter
flashes.

8.

Wait for the end of the measurement with stability control or accept the
calibration value with <ENTER>.
The calibration display for the next buffer appears (voltage display).

9.

If necessary, finish the calibration procedure as a single-point calibra-
tion with <M>.
The calibration record is displayed.

For single-point calibration, the instrument uses the Nernst slope
(-59.2 mV/pH at 25 °C) and determines the zero point of the IDS
pH sensor.

pH

1

-180.0

24.8 °C

01.07.2011

08:00

Buffer

mV

pH

-180.0

24.8 °C

01.07.2011

08:00

AR

Buffer

mV

10. Thoroughly rinse the sensor with deionized water.

11. Immerse the pH sensor in buffer solution 2.