pH value

HandyLab 600

28

ba77064e01

03/2014

Continuing with two-

point calibration

7.

Wait for the measurement with stability control to be completed or ter-
minate the stability control and take over the calibration value with
<ENTER>.
The pH value of the buffer solution is displayed.

8.

Set the nominal buffer value for the measured temperature with
<><>.

9.

Accept the calibration value with <ENTER>.
The calibration display for the next buffer appears (voltage display).

10. If necessary, finish the calibration procedure as a single-point calibration

with <M>.
The calibration record is displayed.

For single-point calibration, the instrument uses the Nernst slope
(-59.2 mV/pH at 25 °C) and determines the zero point of the IDS
pH sensor.

pH

0.0

24.8 °C

ConCal

7.000

01.07.2011

08:00

Buffer

mV

AR

pH

7.000

24.8 °C

ConCal

1

01.07.2011

08:00

Buffer


11. Thoroughly rinse the sensor with deionized water.

12. Immerse the pH sensor in buffer solution 2.

13. Start the measurement with <ENTER>.

The measured value is checked for stability (stability control).
The [AR] status indicator is displayed. The measured parameter
flashes.