2 buffer sets for calibration, Buffer sets for calibration – Xylem HandyLab 600 User Manual

Page 41

Advertising
background image

HandyLab 600

Settings

ba77064e01

03/2014

41

7.1.2 Buffer sets for calibration

You can use the buffer sets quoted in the table for an automatic calibration. The
pH values are valid for the specified temperature values. The temperature
dependence of the pH values is taken into consideration during calibration.

pH resolution

0.001
0.01
0.1

Resolution of the pH display

mV resolution

0.1
1

Resolution of the mV display

Reset

-

Resets all sensor settings to the delivery condition (see sec-
tion 7.4.1 R

ESETTING

THE

MEASUREMENT

SETTINGS

, page 47).

Menu item

Possible
setting
Setting

Explanation

No.

Buffer set *

pH values

at

1

VariCal

Any

Any

2

NIST/DIN
DIN buffers according to DIN 19266
and NIST Traceable Buffers

1.679
4.006
6.865
9.180
12.454

25 °C

3

TEC
Technical buffers

2.000
4.010
7.000
10.011

25 °C

4

Merck 1*

4.000
7.000
9.000

20 °C

5

Merck 2 *

1.000
6.000
8.000
13.000

20 °C

6

Merck 3 *

4.660
6.880
9.220

20 °C

7

Merck 4 *

2.000
4.000
7.000
10.000

20 °C

8

Merck 5 *

4.010
7.000
10.000

25 °C