6 orp, 1 measuring, 1 measuring the orp – Xylem HandyLab 600 User Manual

Page 37: Measuring, Measuring the orp, 6orp

Advertising
background image

HandyLab 600

ORP

ba77064e01

03/2014

37

6

ORP

6.1

Measuring

6.1.1 Measuring the ORP

NOTE
When connecting a grounded PC/printer, measurements cannot be
performed in grounded media as the values would be incorrect. The USB
interface is not galvanically isolated.

Stability control

(AutoRead)

& HOLD function

The stability control function (AutoRead) continually checks the stability of the
measurement signal. The stability has a considerable impact on the reproduc-
ibility of measured values.
The measured parameter flashes on the display

as soon as the measured value is outside the stability range

when the automatic Stability control is switched off.

You can start the Stability control manually at any time, irrespective of the set-
ting for automatic Stability control (see section 7.3.3 A

UTOMATIC

S

TABILITY

CON

-

TROL

, page 46) in the System menu.

IDS ORP sensors are not calibrated. However, you can check IDS
ORP sensors using a test solution.

1.

Connect the ORP sensor to the meter.
The ORP measuring window is displayed.

2.

If necessary, check the ORP sensor using a check solution.

3.

Immerse the ORP sensor in the test sample.

U

157.0

mV

24.8 °C

01.02.2014

08:00

Info

USB output

1.

Freeze the measured value with <AR>.
The [HOLD] status indicator is displayed.
The HOLD function is active.