2 automatic data storage at intervals, Automatic data storage at intervals, Section 8.2 a – Xylem HandyLab 600 User Manual

Page 50: Utomatic, Data, Stor, Intervals

Advertising
background image

Data memory

HandyLab 600

50

ba77064e01

03/2014

8.2

Automatic data storage at intervals

The storage interval (Interval) determines the time interval between automatic
data storage processes. Each data storage process transmits the current data-
set to the interface at the same time.

Configuring the

automatic storage

function

Settings

You can configure the automatic data storage function with the following set-
tings:

1.

Press the <STO_> key.
The menu for automatic data storage appears.

1 Specified entire storage duration
2 Max. available storage duration
3 Graphical display of the memory usage

ID number

1

Interval 30

s

Duration 180

min

continue

0d03h00min

0

1d17h33min

Automatic data storage

1

2

3

01.02.2014

08:00

Back

Menu item

Possible setting

Explanation

ID number

1 ... 10000

ID number for the dataset series.

Interval

1 s, 5 s, 10 s,
30 s, 1 min,
5 min, 10 min,
15 min, 30 min,
60 min

Storage interval.
The lower limit of the storage interval
can be restricted by the number of free
memory locations.
The upper limit is restricted by the stor-
age duration.

Duration

1 min ... x min

Storage duration.
Specifies after which time the auto-
matic data storage should be termi-
nated.
The lower limit of the storage duration
is restricted by the storage interval.
The upper limit is restricted by the
number of free memory locations.

Advertising