7.1.3 Calibration interval

The calibration evaluation is displayed as a sensor symbol.

After the QSC function has been enabled the sensor symbol is replaced by the
QSC scale (see section 5.2.8 QSC

FUNCTION

(

SENSOR

QUALITY

CONTROL

), page

34).

After the specified calibration interval has expired the sensor symbol or the
QSC scale flashes. It is still possible to measure.

Setting the calibra-

tion interval

The calibration interval is set to 7 days in the factory.

21

Reagecon 20 *

2.000
4.000
7.000
10.000
13.000

20 °C

22

Reagecon 25 *

2.000
4.000
7.000
10.000
13.000

25 °C

23

Chemsolute *

2.000
4.000
7.000
10.000

20 °C

24

USABlueBook

*

4.000
7.000
10.000

25 °C

25

YSI *

4.000
7.,000
10.000

25 °C

The buffers are selected in the menu, pH / <ENTER> / Calibration
/ Buffer (see section 7.1.1 S

ETTINGS

FOR

P

H

MEASUREMENTS

, page

40).

To ensure the high measuring accuracy of the measuring system,
calibrate after the calibration interval has expired.