2 general information – Xylem HandyLab 600 User Manual

Page 60

Advertising
background image

What to do if...

HandyLab 600

60

ba77064e01

03/2014

No stable measured

value

Obviously incorrect

measured values

11.2 General information

Sensor symbol

flashes

Display

Cause

Remedy

IDS pH sensor:

– Junction contaminated

– Clean junction

– Membrane contaminated

– Clean membrane

Test sample:

– pH value not stable

– Measure with air excluded if neces-

sary

– Temperature not stable

– Adjust temperature if necessary

IDS pH sensor + test sample:

– Conductivity too low

– Use suitable IDS pH sensor

– Temperature too high

– Use suitable IDS pH sensor

– Organic liquids

– Use suitable IDS pH sensor

Cause

Remedy

IDS pH sensor:

– Sensor unsuitable

– Use suitable IDS sensor

– Temperature difference

between buffer and test sample
too great

– Adjust temperature of buffer or sam-

ple solutions

– Measurement procedure not

suitable

– Follow special procedure

Cause

Remedy

– Calibration interval expired

– Recalibrate the measuring system

Cause

Remedy

– Batteries almost empty

– Replace the batteries (see section

10.1 M

AINTENANCE

, page 57)